Product Manual

Section 10: Performance Verification
10-3
10. Reconnect the FilterLine and verify that the message FilterLine
disappears and the pump is operating.
11. Disconnect the DS-100A - O
XIMAX Durasensor
®
and verify that the
message
SpO
2 Sensor is displayed in the Silent Advisory Message area.
12. Reconnect the DS-100A - O
XIMAX Durasensor
®
and verify that the
message
SpO
2 Sensor disappears.
13. Verify that all buttons on the unit work properly. Refer to the
monitor’s Quick Guide for their applications. (See Figure 10-5:
Quick Guide.)
14. Verify the CO
2 Calibration Check process functions. Follow the
procedure as described below. When completed, return to the
Measuring Mode screen.
Note: Calibration must be performed annually or every 4000 operating
hours , whichever comes first, regardless of the CO
2
Calibration
Check results
Figure 10-5: Quick Guide