Standards and IEC Classifications
a. 1 cycle is at 50 Hz or 20 ms
NOTE: U
T
is the a.c. mains voltage prior to the application of the test level.
Guidance and manufacturer’s declaration - electromagnetic immunity
The Valleylab FT10 FT Series Energy Platform is intended for use in the electromagnetic environment
specified below. The customer or the user of the Valleylab FT10 FT Series Energy Platform should
assure that it is used in such an environment.
Immunity
test
IEC 60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
3 V
rms
150 kHz to 80 MHz
3 V
rms
Portable and mobile RF communications
equipment should be used no closer to
any part of the Valleylab FT10 FT Series
Energy Platform, including cables, than
the recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance:
d=1.2P
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m d=1.2P 80 MHz to 800 MHz
d=2.3P 800 MHz to 2.7 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
a
, should be less than the
compliance level in each frequency
range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol.
Continued
Guidance and manufacturer’s declaration - electromagnetic immunity