Product Manual

Modification and Testing
3-50 Service Manual
Figure3-44.High Modulation and High Light Condition
Verify the following:
a. Active audio alarm.
b. Flashing %SpO2 indication in the range of 73 to 77.
c. BPM indication in the range of 57 to 63.
d. Pulse waveform of approximately 1-inch P-T-P amplitude.
7. Reset the Power On Settings to Factory Defaults. Doing so removes any saved set-
tings and clears the Last Setting option.
8. Turn off the monitoring system.