Operating Manual Filac TM 3000 AD/ADA Electronic Thermometer Table of Contents
Operating Manual Filac™ 3000 AD/ADA Electronic Thermometer
en Table of Contents Section Page I. About Body Temperatures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 II. General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 III. Safety and Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
I. About Body Temperature It is a common myth that 98.6°F (37°C) is the “normal body temperature.” The truth is that 98.6°F (37°C) is an average body temperature. Normal body temperature is actually a range that varies with age, gender, and measurement site. Body temperatures also fluctuate through out the day, typically cooler temperatures in the morning, warmer in the afternoon, and cooling down again in the evening.
III. Safety and Warnings Caution: Federal law (USA) restricts this device to sale by or on the order of a physician. 1. Read this booklet thoroughly before using the Filac 3000 AD/ADA Electronic Thermometer. 2. Do not use this device near flammable anesthetics. Not suitable for use in the presence of flammable anesthetic mixture with air, oxygen or nitrous oxide, or in an oxygen rich environment. 3.
IV.
V. Instructions For Use Probe Covers — Applying & Removing 1. Open probe cover box by lifting tab at top corner and pulling to remove top panel. 2. Insert box of probe covers into top of isolation chamber. (To aid infection control, never switch boxes between blue and red isolation chambers. Also, never switch probes between blue and red isolation chambers. Keep like colors together.) 3. Remove probe from the probe well. This automatically turns on the thermometer. 4.
Oral & Axillary Temperature Taking 1. Make certain that the Blue isolation chamber/probe unit is attached. 2. Withdraw probe and apply a probe cover. The thermometer turns on automatically. 3. An icon identifying Oral or Axillary mode is displayed. The Rectal icon can not be displayed when a blue isolation chamber/probe is attached. 4. Press the Site button on the front panel to select either the Oral or Axillary mode. 5.
Direct Mode 1. The Filac 3000 AD/ADA Electronic Thermometer normally operates in Predictive mode for accurate temperature measuring. In certain cases, such as with a hypothermic patient, the thermometer will automatically shift to Direct (slow) mode and will then act as a temperature monitor. 2. The Filac 3000 AD/ADA Electronic Thermometer can be set to operate exclusively in Direct mode (disable Predictive mode).
Bio-Tech Mode (Option Configuration Menu) 1. To enter Bio-Tech mode, the thermometer must be in sleep mode (probe securely in place with blank screen). 2. Press and hold the Site and °C/°F buttons at the same time for 4 seconds. A “beep” is heard and a dash scrolls in the display. The software version of the device will be shown in the alpha-numeric section of the display 3. The thermometer is now in Bio-Tech mode. To navigate the Bio-Tech mode menu, follow the chart below. 4.
Dashes & Software Version What You See or Hear Direct Mode Lock Option Configuration X, YZ with moving dash DIR with Blinking Anti-Theft (Off, 10, 20,50, 90) THEFT with Blinking Cold Mode COLD with Blinking No Function ON OFF (blank) 10 OFF 20 ON 50 ON 90 ON ON ON OFF ON OFF ON OFF Quick Mode NO FAST with Blinking More Options MORE YES 8 programmable characters, selected alpha-numeric characters blinking Alpha Programming BEEP with Beeper Beeping Continuously Sounds LIGH
VI. Cleaning and Sterilization 1. The entire device may be easily wiped clean. Water temperature should not exceed 130° F (55° C). Do not submerge or soak under water. 2. A mild detergent may be added to water. Use of cleaners such as Spray NineTM*, pHisoHexTM*, HibiclensTM*, Vesta-SydeTM*, or CidexTM* may result in damage to the thermometer case. 3. Use of 10:1 water and hypoclorite mixture or a damp isopropyl alcohol wipe occasionally, is acceptable.
X. Troubleshooting 1. If device fails to function properly after washing & reconnecting probe, rinse all contacts in de-ionized water and blow dry. 2. Icons indicate all other possible errors and remedies. 3.
Water Bath Accuracy (35.5°C/95.9°F to 42.0°C/107.6°F): Direct Mode: Standard Prediction Mode**: Quick Prediction Mode (Oral Only)**: +/-0.1°C - +/- 0.2°F +/-0.1°C - +/- 0.2°F +/-0.3°C - +/- 0.5°F **Greater than 95% of the prediction mode readings will be within the specified accuracy. Patient Accuracy: In standard predict mode, thermometer accuracy meets EN 12470-3 A Standard Prediction Mode reading and a Direct Mode reading will differ by less than ±0.2°C (±0.
XII. Customer Service The circuitry of the Filac 3000 AD/ADA Electronic Thermometer is not customer serviceable. In particular, electronic assembly re-work. will likely affect accuracy. Certain replacement items such as probes and isolation chambers are available from the service centers listed below. All service personnel must be properly trained, qualified and familiar with operation of the thermometer. Improper service may impair proper operation of the Filac 300 AD/ADA Electronic Thermometer.
To Order/Reorder/Replace device components and accessories, specify one or more of the following Reorder Numbers: Probe Covers Case of 500 Case of 2,000 Reorder No. 500500 Reorder No. 502000 Note: Use of any other brand or style probe cover will result in erroneous temperature readings. Complete Oral /Axillary Thermometer Includes electronic thermometer with Blue - Oral/Axillary isolation chamber and Blue - Oral/Axillary probe, with 4-foot (1.
XIII. Warranty Limited Warranty: Covidien warrants to the original purchaser (“Customer”) that this product will be free of defects in materials and workmanship, under normal use, for three (3) years from the date of original purchase from Covidien or its authorized distributor.
Guidance and manufacturer’s declaration – electromagnetic immunity The Filac 3000 Electronic Thermometer is intended for use in the electromagnetic environment specified below. The user of the Filac 3000 Electronic Thermometer should assure that it is used in such an environment.
Guidance and manufacturer’s declaration – electromagnetic immunity The Filac 3000 Electronic Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the Filac 3000 Electronic Thermometer should assure that it is used in such an environment.
Manual No. SMF0212-004 Identification of a substance that is not contained or present within the product or packaging. U.S. Patents (when used with a thermometer probe cover 6,619,837); 6,634,789; 6,839,651; 7,316,507; 7,494,274; 7,549,792; 7,648,268; 7,654,735; D532,710; D535,202; D548,626; D554,008. COVIDIEN, COVIDIEN with logo and Covidien logo are U.S. and internationally registered trademarks of Covidien AG. 0123 TM* Trademark of its respective owner.