Owner's manual

Filac
3000 AD/ADA Electronic Thermometer
11
Table of Contents
Water Bath Accuracy
(35.5°C/95.9°F to 42.0°C/107.6°F):
Direct Mode: +/-0.1°C - +/- 0.2°F
Standard Prediction Mode**: +/-0.1°C - +/- 0.2°F
Quick Prediction Mode (Oral Only)**: +/-0.3°C - +/- 0.5°F
**Greater than 95% of the prediction mode readings will be within the specifiedaccuracy.
Patient Accuracy:
In standard predict mode, thermometer accuracy meets EN 12470-3
A Standard Prediction Mode reading and a Direct Mode reading will differ by
less than ±0.2°C (±0.4°F) on 98% of tested patients
Batteries:
Four“AA”Required.StandardIECpackagesize.Alkaline–1.5Volt
Approximately 6000 temperature readings.
Probe:
Replaceable.
For probe integrity/safety checks, refer to qualified Biomedical Technician or
return toCovidien.
Isolation Chamber:
Removable, replaceable, washable, Polycarbonate/polyester blend.
Device Materials:
Thermometer Case: Flame retardant Polycarbonate/Polyester Blend
Isolation Chamber: Flame retardant Polycarbonate/Polyester Blend
Probe Handle: Flame retardant Polycarbonate/Polyester Blend
Probe Shaft: Flame retardant Polyester
Probe Cable: Polyurethane jacket with TPE over mold
Tip: Aluminum
All materials are latex-free.
Type of Protection Against Electrical Shock:
Internally Powered Equipment. 4 AA Batteries.
Degree of Protection Against Electrical Shock:
Type BF
Degree of Protection Against Ingress of Fluids:
Not Protected-IPX0
Mode of Operation:
Continuous Use . Intended for hand held use.
Degree of Safety of Application in the Presence:
Not suitable for use in the presence of flammable anesthetic mixture with air,
oxygen or nitrous oxide.
Transport and Storage:
Not to exceed 30 days
Transport and store this device between temperatures of
-25°C to 55°C (-13°Fto131°F).
Relative humidity range should be between 10%-95% RH, non-condensing.
Return unit to the factory if the device is dropped and performance appears degraded, or in the event that the environmental
conditions for transport, storage and operation are exceeded.
The Filac 3000 AD/ADA Electronic Thermometer is in compliance with the following International Regulatory and
Safety standards:
UL60601-1 CAN/CSA C22.2 No. 60601-1-08 EN60601-1:2005
EN60601-1-2:2001 EN 12470-3:2000 MDD 93/42/EEC
EN980:2008 ASTM E 1112: 2006