Operating Manual Genius 2 TM Ch ec ke r / For use with Model 303000 Tympanic Thermometers C˚ F/ Ca lib ra to r F/ C˚ Thermometer Checker/Calibrator
en Table of Contents Page Section I - Functional Description.................................................................................................................................................... 1 Section II - Precautions and Warnings............................................................................................................................................ 2 Section III - Icon Identification................................................................................
Section I - Functional Description This manual describes the operation of the Genius™ 2 Checker/Calibrator. This device has been developed to check the accuracy of the Genius™ 2 Tympanic Thermometer and automatically recalibrate the thermometer, if necessary. In addition, the Genius 2 Checker/Calibrator has the capability to generate a test report for each thermometer tested and save it to a USB flash drive. The Genius 2 Checker/Calibrator will work with all Genius™ 2 Thermometer software revisions.
Section II - Precautions and Warnings • The Genius 2 Checker/Calibrator is designed to be operated as specified in this manual. Protection provided by the safety devices in the Genius 2 Checker/Calibrator may be impaired if the device is not used in accordance with the directions contained in this manual. • The Genius 2 Checker/Calibrator enclosure is not waterproof. Avoid spilling liquid of any kind onto the front panel of the device.
RK SYMBOLS 23 Section III - Icon Identification SN CE MARK SYMBOLS Single Use RILE R Serial Number Consult Accompanying Documents 0123 STERILE SingleEO Use RILE A STERILE STERILE R STERILE EO Consult Standby Power CE Mark LOT Dispose of as Electrical and Electronic Waste 40° 0° High Temperature Target Avoid Temperature Extremes F/C˚ FRAGILE 2 STERILIZE NON STERILE Keep Away From Sunlight F/C˚ heric pressure limitation Use By Avoid Temperature Extremes STERILE THIS SIDE UP Doubl
Section IV - Required Equipment 1. 2. 3. 4. 5. 6. 7. Genius 2 Checker/Calibrator Power module with AC adapter Probe cable Operator’s Manual Genius 2 Thermometers to be tested Genius 2 disposable probe covers Small Phillips screwdriver Section V - System Setup The following actions should precede use of the Genius 2 Checker/Calibrator: 1. Connect the correct AC plug adapter for your country to the power module. 2. Connect the power module cable to the power jack on the front panel. 3.
Section VI - Operational Procedure The following procedure will describe the steps required to operate the Genius 2 Checker/Calibrator. In order to facilitate familiarity with the product, it would be advisable to use this section of the manual during the calibration check of the first few thermometers. Installation of USB Flash Drive (Optional) Install a USB flash drive into the USB port located on the front panel if a report file written to the USB flash drive is desired.
Set Date and Time The Set Date and Time screen menu option is used to setup the date and time that is used on the test report. The time and date are internally stored and automatically updated when the unit is powered up. To enter time, use the encoder knob to select the hours and minutes. Note: the hours is setup using a 24 hour clock format, where 1PM is entered as 13:00 etc. When the value is correct, push the encoder knob to move to the next field of the time and then to the date field.
˚ F/ C F/ C ˚ ra to r lib Ca / ke r ec Ch Figure 3 - Proper Insertion of Genius 2 Probe Tip into Target Step 4: The insertion of the probe tip into the target will automatically be detected and the screen display will change to “Press Genius 2 Thermometer scan button.” Press the scan button on the Genius 2 Thermometer while it is inserted into the Low temperature target. The Genius 2 Thermometer will output 3 short beeps as it takes the temperature of the blackbody target.
Step 8: After the pass/fail Results screen has been displayed, press the encoder knob to display a second Results screen containing the target temperatures, the Genius 2 readings, and the target versus thermometer variances. Step 9: The next screen will display “Write report to USB flash drive?” and display Yes and No below. If a report is desired, insert a flash drive into the front panel USB port. Note: allow about 30 seconds after insertion for the USB flash drive to initialize.
Error 7: Wrong Target This error is an indication that the Genius 2 Thermometer has been inserted into wrong target or was inserted into the target prematurely. Please take care to follow the instructions on the screen and not perform steps prior to the instructions being displayed. The Error screen will be displayed for 60 seconds before returning to the Main Menu. To exit the Error screen sooner, press the encoder knob.
General Cleaning Instructions Cleaning of the Genius 2 Checker/Calibrator may be performed as follows: • First, unplug the Genius 2 Checker/Calibrator from power source. Never clean the device while connected to the mains supply. • A mild detergent should be used for general cleaning of outside surfaces. Also, a cloth damped with isopropyl alcohol can be used to clean the unit. Do not use abrasives or solvents. Avoid excess moisture around the target assemblies and the electrical connectors.
Section XI - Customer Service The Genius 2 Checker/Calibrator requires factory calibration and inspection once a year. Notify and return the device to Covidien on or before the calibration date for the Checker/Calibrator. The Checker/Calibrator will display a message one month prior to the calibration date as a reminder. Caution: There are no serviceable parts inside, return to factory for service. In the event that it is necessary to return a unit for repair, please observe the following: 1.
Section XIII - Electromagnetic Conformity Declaration The Genius 2 Checker/Calibrator has been built and tested according to IEC/EN61010-1 and EN61326-1 standards The Checker/Calibrator is intended for use in the electromagnetic environment specified below. The user of the equipment should assure that it is used in such an environment.
Identification of a substance that is not contained or present within the product or packaging. Identification d’une substance non présente dans le produit ou l’emballage. Identifizierung einer Substanz, die weder im Produkt noch in der Verpackung enthalten bzw. vorhanden ist. Identificazione di una sostanza che non è contenuta né presente nel prodotto o nella confezione. Identificación de una sustancia no contenida o que no esté presente dentro del producto o embalaje.