User Manual

Genius
TM
2 Checker/Calibrator
Table of Contents
1
Section I - Functional Description
This manual describes the operation of the Genius™ 2 Checker/Calibrator. This device has been developed to check the accuracy
of the Genius™ 2 Tympanic Thermometer and automatically recalibrate the thermometer, if necessary. In addition, the Genius 2
Checker/Calibrator has the capability to generate a test report for each thermometer tested and save it to a USB ash drive.
The Genius 2 Checker/Calibrator will work with all Genius™ 2 Thermometer software revisions.
The Genius 2 Checker/Calibrator contains two independently controlled infrared calibration targets that are similar to factory
calibration targets. These “blackbody targets are designed to have ecient radiative heat transfer and produce infrared radiation
that approximates the human ear at temperatures of 32.2° C (90.0° F ) and 40.6° C (105.0° F ).
F/C˚
F/C˚
1 2 3 4
8 910
5
6
7
11
F 3.15A 250V
Checker / Calibrator
Figure 1 - Front Panel Layout
Legend
1 Power Connector 7 High Temperature Target
2 Power Switch 8 High Temperature Target LED
3 Fuse Holder 9 LCD Display
4 Low Temperature Target LED 10 Encoder Knob
5 Low Temperature Target 11 USB port
6 Genius 2 Thermometer Connector