Calibrated Accuracy Limits (after Recalibration*):
Ambient Temperature Target Temperature Accuracy
16°C to 33°C 36°C to 39°C ± 0.2°C
(60.8°F to 91.4°F) (96.8°F to 102.2°F) (± 0.4°F)
16°C to 33°C less than 36°C ± 0.3°C
(60.8°F to 91.4°F) (less than 96.8°F) (± 0.5°F)
16°C to 33°C greater than 39°C ± 0.3°C
(60.8°F to 91.4°F) (greater than 102.2°F) (± 0.5°F)
*Post recalibration accuracy using the Genius 2 Checker/Calibrator may not necessarily be equivalent to factory calibration.
Displayed Temperature Measurement Range:
Temperature Range depends on Site Mode as follows:
Mode Range °C Range °F
Ear 33.0 to 42.0 91.4 to 107.6
Oral 33.6 to 42.0 92.5 to 107.6
Core 34.0 to 42.0 93.2 to 107.6
Rectal 34.2 to 42.0 93.6 to 107.6
Ambient Temperature Range:
16°C to 33°C (60.8°F to 91.4°F) where as CEN denes the standard as 16°C to 35°C (60.8°F to 95°F), 10 to 95 %RH, non-condensing.
Storage Temperature Range:
-25°C to 55°C (-13°F to 131°F), up to 95% RH non-condensing. If the unit is stored at extremes, it is recommended that the unit be
checked on the eld calibration checker or at the factory before returning to service.
Clinical Repeatability:
Meets Section A.5 of EN 12470-5: 2003 (E) per Covidien llc technical report. Data is available from Covidien llc onrequest.
Response Time:
Less than 2 seconds
Pulse Timer:
60 seconds
Temperature Resolution:
0.1°C or 0.1°F
Power:
3 AAA alkaline batteries
Battery Life:
Minimum of 15,000 temperature readings
Size:
Thermometer — 17.8 cm (7”)
Base — 20.3 cm (8”)
Weight:
Thermometer (with batteries) — 160 grams
Base — 100 grams