Nellcor™ Bedside SpO2 Patient Monitoring System, PM100N In-Service Presentation COVIDIEN, COVIDIEN with logo, Covidien logo and positive results for life are U.S. and internationally registered trademarks of Covidien AG. Other brands are trademarks of a Covidien company. ©2015 Covidien.
The Nellcor™ Bedside SpO2 Patient Monitoring System, PM100N, is intended for prescription use only for continuous monitoring of functional arterial oxygen saturation and pulse rate of adult, pediatric and neonatal patients, who are well or poorly perfused, during motion and no motion conditions. It is intended to be used in hospital, hospital type facilities, and intra-hospital transport as well as in the homecare environments.
New Homecare and Sleep Study Modes Nellcor™ Bedside SpO2 Patient Monitoring System, PM100N is the only bedside monitor with standard, homecare, and sleep study modes. • Standard Mode: Advanced functionality to monitor in-hospital patients with SpO2, pulse rate, and SatSeconds alarm management. • Homecare Mode: Password protected, simplified monitoring for non-hospital settings. Prevents lay users from modifying alarm settings. • Sleep Study Mode: Password protected, simplified monitoring for sleep studies.
Rear Panel 1. Nurse Call Port 2. AC Power Connector 3.
Front & Side Panel Audio Pause Silence audible alarms for set time Home Button Return to this main screen Power Button Press and HOLD to turn on and off USB Ports For data download and upgrades SpO2 Connector Insert interface cable here Knob Rotate and press, to control system functions
Display – Right Side AC Power Indicator Battery Charge Indicator Interference Indicator Sensor Disconnect Indicator Sensor Off Indicator Sensor Message Indicator
Hands on Review • Attach Power Cord to device and plug into AC receptacle • Power on device and determine when it is ready for use • Connect cable and sensor to device • Create the various sensor indicator messages – – – – Interference indicator Sensor disconnect indicator Sensor off indicator Sensor message indicator (articulate when this occurs) • Power off the device
Standard Monitoring Display – Main View SpO2: Upper/Lower Alarm Limits Pulse Rate: Real Time Value Battery Indicator SpO2: Real Time Value Plethysmographic Waveform (Pleth) Blip Bar Informative Message Area Pulse Rate: Upper/Lower Alarm Limits
Standard Monitoring Display – Bottom Informative Message Area Patient Mode Alarm Limits Menu Options Menu 9 Covidien Respiratory & Monitoring Solutions | 4/10/15 |
Standard Monitoring Alarm Limits Menu To access the Alarm Limits Menu, rotate the knob to highlight the alarm limits display area and press.
Standard Monitoring Quick Access Menus For quick access to alarm settings, rotate the knob to highlight one of two areas and press.
Hands on Review • Attach the sensor to a patient – Determine the patient’s real time SpO2 and PR reading – Set parameter alarm limits • SpO2 target range: 85% to 100% • Pulse rate target range: 80 bpm to 110 bpm • Once alarm limits are set – Discuss alarm status • What do you see? • What do you hear? – How would you respond to the alarm(s)? Covidien Respiratory & Monitoring Solutions | 4/10/15 |
Standard Monitoring Patient Mode Menu To access the Patient Mode Menu, turn the knob to highlight the patient mode display area and press the knob.
Standard Monitoring Options Menu To access the Options Menu, turn the knob to highlight the Options Icon and press the knob. Press as a shortcut to access the Options Menu Volume Adjustment • Alarm • Key Beep • Pulse Patient Response Mode • Normal • Fast Clear Trend Data Service Menu 1. 2. 3. 4. 5. Power On Settings Alarm Audio Paused Alarm Audio Reminder Permission to Mute Alarm Language 6. 7. 8. 9. Date/Time Setting System Information System Test Trend Data Download Settings 10. 11. 12. 13.
Hands on Review • Your next patient is a 1 year old child. – Set the pulse oximeter to the appropriate patient mode. • It is very loud in the patient’s room. – Set the alarm volume to the loudest setting for SpO2 and PR.
Standard Monitoring SpO2 Waveform Menu To access the SpO2 Waveform Menu, turn the knob to highlight the waveform or trend display area and press the knob. Sweep Speed • 6.25mm/s • 12.
Standard Monitoring Sweep Speed Overview Allows the clinician to set the speed to where the SpO2 waveform trace moves across the screen. Sweep Speed = 25.0mm/s (i.e. “Zoom In”) The higher the sweep speed value, the more granular the data appears on the screen. Sweep Speed = 6.25mm/s (i.e.
Standard Monitoring Tabular Trend View To access the Tabular Trend View, turn the knob to highlight the waveform or trend display area and press the knob. Select Tabular Trend and press the knob.
Standard Monitoring Tabular Trend Scroll To scroll through the Tabular Trend View, rotate the knob to the left for historical data and to the right to get back to present time Scroll Interval Options • • • • 1 (every 4 sec) 5 (every 20 sec) 100 (~6.5 min) 500 (~33.5 min) Turn knob to the left for historic data Note: To change Scroll highlight Interval Options, press the knob.
Standard Monitoring Trend Data Download To access the Trend Data Download Menu, turn the knob to highlight the Options Menu and press. Mini-USB Port 1. Connect mini-USB port to a PC 2. Select Options 3. Select Trend Data Download 4.
Standard Monitoring Graphical Trend View To access the Graphical Trend View, turn the knob to highlight the waveform or trend display area and press the knob. Select Graphical Trend and press the knob.
Hands on Review • Access tabular trend data for your patient – How have your patient’s SpO2 values changed in the last hour? – How have your patient’s PR values changed in the last hour? • Use the scroll interval options to review your patient’s data from 3 hours ago. • Download your patient’s data using the data download capability.
Standard Monitoring SatSeconds™ Alarm Management Setting SatSeconds Setting SatSeconds Icon
Standard Monitoring Alarm Limits Menu To access SatSeconds™ Alarm Limits Menu, turn the knob to highlight the alarm limits display area and press the knob.
Hands on Review • Activate SatSeconds™ Alarm Management – If you want to minimize clinically insignificant alarms that occur with your patient, set SatSeconds to 100.
Homecare Mode To access Homecare Mode, turn the knob to highlight and select the Options Menu. Within menu, select the Change Patient Mode display area and press the knob.
Homecare Mode After selecting Change Patient Mode display area, highlight Switch to Homecare and press the knob. Turn the knob to highlight Switch to Homecare and the press knob.
Homecare Mode After selecting Switch to Homecare display area, enter passcode. Entering Passcode: • Turn the knob to highlight the number in passcode and press the knob. • Once number is highlighted in yellow, turn knob to change number then press the knob to select. • Once number is selected, repeat process for remaining passcode numbers.
Homecare Mode After passcode is entered, a confirmation screen will pop up. Select OK and then you will be in Homecare Mode.
Sleep Study Mode To access Sleep Study Mode, turn the knob to highlight the Options Menu and select the Change Patient Mode display area and press the knob.
Sleep Study Mode After selecting the Change Patient Mode , then highlight Switch to Sleep Study and press the knob.