Owner`s manual

23
REPLACING THE FUSES
WARNING: To avoid electric shock, disconnect the test leads
from any source of voltage before removing the fuse cover.
1. Disconnect the test leads from the meter.
2. Remove the protective rubber holster.
3. Remove the Phillips head screw located on the lower
back of the instrument.
4. Remove the fuse/battery compartment cover to access
the fuses.
5. Gently remove the fuse(s) and install new fuse(s) into
the holder(s).
6. Always use fuses of the proper size and value
(200mA/250V fast blow for the mA / !A ranges,
10A/250V fast blow for the A range).
7. Secure the fuse/battery compartment cover.
8. Place the protective rubber holster on the meter.
WARNING: To avoid electric shock, do not operate your meter
until the fuse cover is in place and fastened securely.
1
2
3
4
5
6