Operation Manual
62 Electromagnetic compatibility
GB IRE
Recommended separation distances
Recommended separation distances between portable and mobile RF communications
equipment and the EQUIPMENT or SYSTEM – for EQUIPMENT or SYSTEM that are not LIFE-
SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and the Sphygmomano-
meter (MD4000)
The Sphygmomanometer (MD4000) is intended for use in an electromagnetic environment in which radiated RF disturban-
ces are controlled. The customer or the user of the Sphygmomanometer (MD4000) can help prevent electromagnetic interfe-
rence by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and
the Sphygmomanometer (MD4000) as recommended below, according to the maximum output power of the communica-
tions equipment.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter (m)
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0,01 0.1167 0.2334
0,1 0.3689 0.7378
1 1.1667 2.3334
10 3.6893 7. 3 7 8 6
100 11.6667 23.3334