compartment temperature and maintain Class 100 integrity. If you are
unable to exhaust the minimum 30 CFM from the control compartment,
then you must limit the maximum operating temperature to 200°C.
Operation of the unit above 200°C, without moving the required 30 CFM
through the control compartments exhaust, will result in control
degradation and process failure. Not supplying the required 30 CFM may
also keep the unit from operating at class 100 specifications by allowing
particulars from the control compartment to escape into the room.