Instruction manual
11 
OPERATION 
All control functions are covered in the Jumo Manual. This is a simplified 
version. 
Check power supply against unit serial plate. They must match. 
On ovens that need to be hard wire, must be done by a qualified electrical 
technician. 
7.1  Push the power switch to the I/ON position, turn the high limit thermostat 
to its maximum position clockwise using a flat-head screw driver. The 
main temperature controller should be illuminated. The top display is the 
Process Temperature and the bottom display is the Set Point. 
7.2  AUTO TUNING JUMO CONTROL FOR OPTIMUM CONTROL 
Even though the control was Auto Tuned at the factory, you might choose 
to tune the control to your specific Set Point. 
1.  Push the down arrow once.   
2.  Once the down arrow has been pushed, use the up and down 
arrow  to achieve desired set point. 
3.  After reaching desired set point, the display should blink once to enter 
the value. 
4.  To Auto Tune, push the up and down arrow 
 simultaneously for 2 
seconds until the Set Point Display flashes Tune. 
5.  After  Tuning Cycle is complete, the bottom display will stop flashing 
Tune ( ) and return to desired Set Point. 
7.3 CHANGING SET POINT ON CONTROL WHEN SETUP AS 
PROGRAMABLE CONTROLLER 
1.  To change Set Point, push the 
 Down Arrow once and then use the 
 Up and Down Arrow to change the Setpoint. 
CAUTION: If the 
 Up Arrow is pushed first, it will start a program and 
will not allow you to change the Setpoint. To return to NORMAL 
DISPLAY, push the   Up Arrow again. 
7.4  CHANGING CONTROL FUNCTION TO SINLE SET POINT CONTROL 
1.  Push the PGM Button once and then navigate using the   Up and 
Down Arrows until you reach “CONF” in the bottom display. 
Section 
7