Instruction manual
13 
and a RAMP/PROGRAM symbol   will appear in the lower right side of 
the display. This indicates the program is running. To stop the program, 
push the up arrow   again and the RAMP/PROGRAM symbol   will 
disappear and the program will end. After program ends, control returns 
to previous Set Point (starting Set Point). This is explained in Section 6.5 
Paragraph 4. 
OFFSET CALIBRATION 
This is used to match the process temperature to an independent 
thermometer. The Offset Calibration is measured in one degree 
increments. For example, if the process value reads 100°C and the 
independent thermometer reads 98°C, the Offset Calibration would be set 
at -2 to match the process value to the independent thermometer. To 
navigate the control to the Offset Calibration mode, push the PGM button 
 once and then push the up arrow   once and the bottom display 
should read CONF. Then push the PGM button again and the display 
should say INP. Push the PGM button   again and the display should 
read INP1. Push the PGM button   again and the display should read 
SENS. This enters you in the Input/Configuration Menu. To navigate 
through the menu, press the up and down arrows.   Press the up or 
down arrow   repeatedly until you reach OFFS in the bottom display.  
Then, push the PGM button   once, and the display will start flashing.  
While flashing, use the up or down arrow   to set the desired Offset.  
After the display stops flashing, push the Exit Button   repeatedly until 
the NORMAL Display is reached. 
7.5  Setting High Limit: Prior to setting the High Limit Control be certain the 
Main temperature Control has reached the input set point and has 
remained stable for several hours. This will insure that the High Limit is 
set at the proper temperature. As stated in 4.3, the High Limit control 
should be at maximum position, clockwise. Now turn the thermostat 
counter-clockwise using a flat-head screwdriver until the High Limit 
Activated indicator light comes on. Next turn the thermostat clockwise just 
until the indicator light turns off. Then turn the thermostat clockwise two 
(2) minor increments on its scale past the point where the indicator light 
went out. This will set the High Limit at approximately 10°C above the 
Main Temperature set point. 