Instruction manual
14 
MAINTENANCE 
8.1 Cleaning:  The unit chamber should be cleaned and disinfected on a 
regular basis. Remove shelving & shelving supports then clean 
thoroughly, including all corners using a suitable disinfectant that is 
appropriate to your application. Regular periodic cleaning is required.  
Special care should be taken when cleaning around sensing heads to 
prevent damage. DO NOT use chlorine based bleaches or abrasive 
cleaners as this will damage the stainless steel interior. DO NOT use 
spray cleaners that might leak through openings and cracks and get on 
electrical parts or that may contain solvents that will harm the coatings. 
Warning: Never clean the unit with alcohol or flammable cleaners with 
the unit connected to the electrical supply. Always disconnect the unit 
from the electrical service when cleaning and assure all volatile or 
flammable cleaners are evaporated and dry before reattaching the unit to 
the power supply. 
8.2 Storage:  If the oven is to be turned off it can be reactivated without 
controller adjustments. Prior to storage the interior should be wiped dry to 
eliminate contamination. If the unit is to be transported disconnect the 
power supply, remove shelving and screw the leveling feet in. See 
Section 3.3 Lifting/handling for transport instructions. 
8.3  There is No Maintenance required on the electrical components. If the 
unit fails to operate as specified please review the troubleshooting guide 
prior to calling for technical support. 
8.4  Hepa Filter Replacement: Replace the hepa filter when the control 
compartment exhaust becomes noticeably reduced. The amount of time a 
Hepa filter will last depends on the environment in which the unit operates. 
This will have to be determined by the end user to establish standard 
operating procedures or maintenance schedule. 
Section 
8