Instruction manual
15 
TROUBLESHOOTING 
TEMPERATURE 
Temperature too high 
1/ controller set too high-see your Jumo users manual 
2/ controller failed on – call Customer Service 
3/ wiring error – call Customer Service 
Display reads "HI" or "400"+ 
Probe is unplugged, is broken or wire to sensor is broken – 
trace wire from display to probe; move wire and watch display 
to see intermittent problems 
Chamber temp spikes over set point 
and then settles to set point 
Recalibrate – see Jumo users manual 
Temperature too low 
1/ high limit set too low – see section 4.5 
2/ controller set too low – see Jumo users manual 
3/ unit not recovered from door opening – wait for display to 
stop changing 
4/ unit not recovered from power failure or being turned off – 
incubators will need 24 hours to warm up and stabilize 
5/ element failure – call Customer Service 
6/ controller failure – confirm with front panel lights that 
controller is calling for heat 
7/ high limit failure – confirm with front panel lights that High 
Limit is operating correctly 
8/ wiring problem – check all functions and compare wiring to 
owners manual - especially around any areas recently worked 
on 
9/ loose connection – check shadow box for loose 
connections 
Display reads "LO" 
  1/ sensor is plugged in backwards – call Customer Service 
2/ if ambient room temperature is lower than range of unit – 
compare set points and ambient temperature to rated 
specifications in Section 10.0 
Unit will not heat over a temperature 
that is below set point 
1/ confirm that blower motor is moving and that amperage 
and voltage match data plate – check blower motor motion in 
shadow box and feel for air movement in chamber 
2/ confirm that set point is set high enough –turn High Limit 
clockwise and see if High Limit light comes on 
3 /  check connections to sensor 
4/ check calibration – using independent reference 
thermometer, follow instructions in Jumo users manual  
Section 
9










