User manual

Table Of Contents
An example of the initialization sequence is shown in Figures 7-58 and 7-59 with
automated CTD and ACM verification, and a Sea-Bird 52MP CTD.
Step
í
Step
î
Step
ï
McLane Research Laboratories, USA
McLane Moored Profiler
Version: MMP-4_01 S/N: ML12345-00D
╔═════════════════════════════════╗
Main Menu
╚═════════════════════════════════╝
Wed Aug 2 12:06:04 2006
<1> Set Time <5> Bench Test
<2> Diagnostics <6> Deploy Profiler
<3> Flash Card Ops <7> Offload Deployment Data
<4> Sleep <8> Contacting McLane
<C> Configure
Selection ? 6
Clock reads 08/02/2006 12:06:52
Change time & date (Yes/No) [N] ? n
Setting watchdog clock . . . done.
Profiling history:
Total motor hours: 15.33 hours
Total meters traveled: 8456 meters
Verification of sensor settings
<A> Automated (standard settings only)
<M> Manual (operator controlled)
<S> Skip Verification
Selection ? a
Automated verification of sensor settings.
Applying power to CTD . . .
Starting verification . . .
Sending command ***O
Open Mode
Expected response received.
Proceeding to next command.
Sending command ROP
1363
Scale is OFF
Address Operation is OFF
Auto is ON
Expected response received.
Standard CTD settings verified.
Figure 7-58: Deployment Initialization (screen 1 of 2)
7-38