User manual

Table Of Contents
Programming a Deployment
When deployment initialization is complete, the Deployment Menu displays. The menu
re-displays after each change.
ID M| Mooring ID = 001
Start Z| Countdown delay = 00:05:00 [HH:MM:SS]
Schedule I| Profile start interval = 000 00:10:00 [DDD HH:MM:SS]
R| Reference date/time = 03/21/2007 11:30:00
B| Burst Interval = Disabled
N| Profiles per burst = Disabled
P| Paired profiles Disabled
F| Profiles / file set = 1
Stops S| Shallow pressure = 2.5 [dbar]
D| Deep pressure = 12.0 [dbar]
H| Shallow error = 1.0 [dbar]
E| Deep error = 1.0 [dbar]
T| Profile time limit = 00:02:00 [HH:MM:SS]
C| Stop check interval = 2 [sec]
L| Fluorometer = Disabled
O| OBS Turbidity = Disabled
Endurance | Power for single profile = 6.1 [mAh]
| Total profiles/(240 Ah) = 39144
| Est. battery expiration = 12/18/2007 07:46:47
Deploy V| Verify and Proceed
Figure 7-62: Deployment Menu (firmware version 4.04)
Deployment parameters are grouped on the display by general function: Mooring ID,
Start, Schedule, Stops, Endurance and Deploy. The Start parameters (countdown delay or
specific date and time) control the start time of Profile 0 (Dive 0). The MMP executes Profile 0
to move to the bottom stop. The Start setting provides time to launch the MMP and set the
mooring before profiling begins.
A Quick Reference overview of each deployment parameter is provided next. More
detailed descriptions of the deployment parameters follow the Quick Reference.
7-42