Profile Start Interval/Pair Start Interval <I>
Profile start interval is the time between profiles (or pairs). If the interval is less than the
time required to complete a profile (or pairs), the next profile (or pair) will be skipped to prevent
asynchronous profiling. Enter a Profile Start Interval greater than the Profile Time Limit.
Profile start
interval
Profile time
limit
Schedule I| Profile start interval = 000 01:30:00 [DDD HH:MM:SS]
R| Reference date/time = 11/03/2006 12:00:00
B| Burst interval = 010 00:00:00 [DDD HH:MM:SS]
N| Profiles per burst = 4
P| Paired profiles Disabled
F| Profiles / file set = 1
Stops S| Shallow pressure = 50.0 [dbar]
D| Deep pressure = 1050.0 [dbar]
H| Shallow error = 500.0 [dbar]
E| Deep error = 500.0 [dbar]
T| Profile time limit = 01:23:20 [HH:MM:SS]
C| Stop check interval = 30 [sec]
L| Fluorometer = Disabled
O| OBS Turbidity = Disabled
Endurance | Power for single profile = 301.9 [mAh]
| Total profiles/(240 Ah) = 712
| Est. battery expiration = 09/18/2011 08:00:00
Deploy V| Verify and Proceed
Selection ? i
Enter profile start interval
Days ( 0 to 366) ? 0
Hours ( 0 to 23) ? 1
Minutes ( 0 to 59) ? 30
Seconds ( 0 to 59) ? 0
Figure 7-63: <I> Profile Start Interval
Figure 7-64 shows a deployment with single profiles.
Profile Interval
Profiles
Top
Bottom
00
Start
Time
Reference
Time
Reference
Time
Dive 0
22
11
Time
Figure 7-64: Profiles Example