The MMP continues a burst until it has completed all of the profiles or pairs in the burst.
If this requires longer than the burst interval, the next burst (or bursts) will be skipped. Range is
1 1000 profiles or pairs of profiles. Figure 7-67 shows a deployment with bursts.
00
Dive 0
Bursts
Reference
Time
Top
Bottom
22 3311 44
Profile
Interval
Start
Time
55 66
Burst 1
Time
Burst Interval
Figure 7-67: Burst Example
Paired Profiles <P>
When Paired Profiles are enabled, the deployment is scheduled in up-down pairs. The
down profile starts as soon as the up profile data storage is completed. The start interval is the
time between the start of motion for sequential up profiles. When paired profiles are disabled, the
deployment is conducted in single profiles and the start interval is the time between the start of
motion for sequential profiles. Figure 7-68 shows a deployment with paired profiles.
Pairs
Top
Bottom
00
Start
Time
Dive 0
Time
2211
Pair Interval
Reference
Time
Reference
Time
33 44
Figure 7-68: Pairs Example
Figure 7-69 shown next illustrates three deployment programming styles, for visual
comparison.
Appendix 7-49