<L> Fluorometer
The optional Seapoint Fluorometer analog sensor can be set to automatic gain or one of 4
fixed gain levels. The average number of samples can also be operator-defined. If the
Fluorometer is enabled, these settings can be changed (the Fluorometer sensor can be enabled
only from the Configuration menu). For more information, see the ‘Optional Seapoint Analog
Sensors’ appendix in this User Manual.
<O> OBS Turbidity
The optional Seapoint Turbidity analog sensor can be set to an automatic gain or one of 4
fixed gain levels. The average number of samples can also be operator-defined. If Turbidity is
enabled, these settings can be changed (the Turbidity sensor can be enabled only from the
Configuration menu). For more information, see the ‘Optional Seapoint Analog Sensors’
appendix in this User Manual.
Endurance Parameters
Endurance parameters show an estimated battery expiration. The estimate adjusts each
time a deployment parameter is changed. The single profile battery drain and total profiles
/battery are also listed.
Single Profile Current
An estimate in mAh of the battery drain for a single profile.
Total Profiles/(Ah)
The estimated total profiles per battery pack. The default battery energy is 240Ah (the
energy of the L24-1000 battery). Change this default, if necessary in the ‘Nominal Endurance’
option on the System Configuration menu.
Est. Battery Expiration
An estimate of battery expiration based on battery energy (the default is 240Ah), the
single profile drain, and the total number of profiles. The estimate is updated each time a
deployment parameter is changed.
Deploy
<V> Verify and Proceed
This option indicates that deployment parameter selection is complete. The firmware
conducts parameter range and consistency checks and estimates battery expiration. Range
violations and parameter inconsistencies trigger user prompts for changes.