User manual

Table Of Contents
IMPORTANT
Archive AUTOEXEC.BAT and MMP-N_NN.RUN onto a PC hard drive and bring copies to
every deployment. If either of these files is inadvertently removed from the flash card the MMP
will be inoperable. A backup of each program is provided with the MMP.
To install AUTOEXEC.BAT and the MMP-N_NN.RUN backup programs that are
included with the MMP, insert the flash card to a PC and copy the files from the flash card to the
PC using standard DOS or Windows commands.
IMPORTANT
Before copying the files, archive all data files, create a backup and then reformat the flash card
to avoid fragmenting the file structure on the card.
When the system switches on, the screen displays the total data files that the flash card
can accommodate and the maximum file sets that can be saved.
File system can accommodate 4091 data files
Deployment can include up to 1361 file sets
The DOS file system limits the number of files that can be stored in the root directory of
the flash card to 4096. The system files stored on the flashcard (for example, AUTOEXEC.BAT,
DEPLOY.DAT, and so on) are subtracted from 4096 to determine the number of data files that
the file system can accommodate.
The ‘Profiles/file set’ option allows more profiles to be conducted by storing multiple
profiles as a single data file. The maximum profiles per file set is affected by the configuration of
the instrument. The 1361 file set maximum is based on an MMP with both a CTD and an ACM
installed which generates 3 files for each deployment.
The Profiles/file set option is described in Chapter 7, “MMP Firmware 4.X User
Interface”.
Appendix B-5