User manual

Table Of Contents
sampling will prevent the profiler from spending a significant period of time parked at the top of
the range.
Deployment parameters for the first iteration, meeting all of the proposed requirements,
are shown below.
Start 1| Countdown delay = 12:00:00 [HH:MM:SS]
5| Pairs per burst = Disabled
Selection ? d
Evaluate the
schedule
More than 13
hours are
needed to
complete a pair
of profiles
Continue
checks
Schedule 2| Pair start interval = 000 12:00:00 [DDD HH:MM:SS]
3| Reference date/time = 04/10/2002 00:00:00
4| Burst interval = 007 00:00:00 [DDD HH:MM:SS]
6| Paired profiles Enabled
Stops 7| Shallow pressure = 50.0 [dbar]
8| Deep pressure = 5500.0 [dbar]
9| Shallow error = 250.0 [dbar]
A| Deep error = 50.0 [dbar]
B| Profile time limit = 07:34:10 [HH:MM:SS]
C| Stop check interval = 30 [sec]
D| Done
Checking entries . . . done.
All entries are within allowed ranges.
Checking profile schedule . . . done.
Calculated minimum profile pair duration: 13:28:08
Estimated minimum profile pair duration exceeds pair
start interval. System will skip pairs whenever a
pair start time is missed.
Change selections (Yes/No) [Y] ? n
Checking burst schedule . . . done.
Burst mode disabled.
Profiling distance will reach 1 million meters ,
Estimated Profile 1 start time: 04/10/2002 00:00:00
on 07/10/2002 after 183 profiles.
Accept and store schedule selections (Yes/No) [Y] ? n
Figure C-20: Deep Water Deployment Parameters- First Settings
Appendix C-21