User manual

Table Of Contents
Powering Down the MMP
To power off the MMP, complete the following steps:
4. Return to the Main Menu.
5. Select ‘Sleep’ from the menu.
6. Disconnect the main lithium battery pack.
IMPORTANT
Do not disconnect power to stop a deployment. Powering down during deployment may corrupt
open data files.
The Main Menu - Operating the MMP
The MMP Main Menu (shown below) controls all system operations. This section of the
manual describes each of the visible (and hidden) Main Menu options. To select an option type
the appropriate alphanumeric character key and press [ENTER].
McLane Research Laboratories, USA
McLane Moored Profiler
Version: MMP-3_12.c S/N: ML11827-01C
╔═════════════════════════════════╗
Main Menu
╚═════════════════════════════════╝
Wed Dec 29 12:02:53 2004
<1> Set Time <5> Bench Test
<2> Diagnostics <6> Deploy Profiler
<3> Flash Card Ops <7> Offload Deployment Data
<4> Sleep <8> Contacting McLane
Selection ?
Figure G-2: Main Menu (version 3.12)
<1> Set Time
This option allows the operator to set the real time clock (RTC). The system
automatically sets the watchdog circuit (WDC) whenever the operator sets the RTC.
G-5