13.7841, 21.7325, 0.4280
TX channel closed.
Shutting down power to CTD.
Press [ENTER]
and CTD
responds with
scan
Continue to
press [ENTER]
for additional
scans
Type [CTRL]-
[C] and system
puts CTD in
open mode,
shuts down
communication
channels, and
shuts down
CTD
Selection ? 1
Applying power and enabling COMMs to CTD . . . ready.
Initializing TT8 communication channels . . .
TX channel open.
RX channel open.
Hailing frequencies open.
<CTRL>-<C> to terminate session.
13.7847, 21.7239, 0.2079
13.7845, 21.7330, 0.4481
RX channel closed.
Press any key to continue.
Figure G-16: Triggering CTD to Send Data Scan
Each scan displays conductivity (mmho/cm), temperature (°C) and pressure (dbar).
Open Mode
Type ‘O’
[ENTER] to
place CTD in
open mode
***O
Figure G-17: CTD in Open Mode
IMPORTANT
To terminate the communication session and power down the sensor, press [CTRL]-[C] at
anytime, regardless of the current operational mode of the CTD. Note that if you are operating
the MMP with Version 2.07 or earlier you must place the CTD (and ACM) in open mode
manually before terminating the session. Some early models of the CTD could lose calibration
constants if shut down in run mode. See the CTD manual for further information.
Standard CTD settings exist for use with the MMP. The system will check for these
settings at the discretion of the operator before beginning a deployment. However, the system
will not automatically program the standard settings. The CTD pass-through utility can be used
to verify and program the standard settings before a deployment.
G-14