User manual

Table Of Contents
Verifying CTD Settings
To use the CTD pass-through utility, complete the following steps:
11. From the Bench Test option on the Main Menu, select <1> CTD Communications.
12. Type ***O [ENTER] to place the CTD in open mode.
13. Type ROP [ENTER]. The system should display the following:
serial number 1335
Scale is OFF
Address Operation is OFF
Auto is ON
Figure G-18: CTD Settings
14. If the response to Read Operational Parameters (ROP), varies from the expected result,
enter commands from the list below to correct the deviations:
Command Result
SCALE=OFF[ENTER]
Clears the scale
CAOP[ENTER]
Clears the address operation
AUTO=ON[ENTER]
Sets auto
15. Type ROP[ENTER] to verify the changes.
16. Type ***E[ENTER] to store settings in the CTD EEPROM.
17. Wait a few seconds for the CTD to send a carriage return to the screen, indicating that the
storage operation is complete.
18. Press [ENTER] to return the CTD to Open Mode.
19. Press [CTRL]-[C] to terminate the session and shut down the CTD.
20. Re-establish communications and verify that the settings are now correct on power-up.
A sample screen for verifying standard CTD settings is shown next.
G-15