Pressure = -0.712 dbar
Press any key to continue.
Figure G-20: Testing Pressure Information
Option <3> CTD Average Pressure
This option can be used to acquire a time series of the power-up response of the CTD
pressure transducer.
NOTE (Firmware Version 2.07)
If you are operating a system with a version prior to 2.08, upgrade before using this option.
Versions prior to 2.08 shut down the CTD in run mode, which is not optimal.
This option queries the operator for the number of pressure values to average, starts the
CTD, queries the requested number of scans, parses the responses, displays the results, calculates
and displays the average, and shuts down the CTD. An example is shown next.
Operator
requests 3
measurements
Enter number of measurements for average (1 to 5000) ? 3
Applying power to CTD . . . ready.
-0.523 dbars
+0.015 dbars
+0.222 dbars
Average pressure = -0.095 dbar
Press any key to continue.
Figure G-21: CTD Average Pressure