User manual

Table Of Contents
number of bytes per scan in the compressed binary format.
NAME
[ENTER] displays the
status of that particular feature.
19. Type ROP [ENTER]. The system should display the following:
Continuous clear
address op clear
Logging Ops Set
RLD checksum output cleared
Figure G-26: ACM Operational Parameters
If the response to Read Operational Parameters (ROP), varies from the expected one, enter
commands from the list below to correct the settings:
Command Result
CCOP[ENTER]
Clears continuous
CAOP[ENTER]
Clears address operation
SLOG[ENTER]
Sets logging ops (the operator cannot control the checksum output)
TILT[ENTER]
TILT= on
(Enables tilt)
COMP[ENTER]
COMP = on
(Enables compass)
NRML[ENTER]
NRML = on
(Sets compass normalization)
20. If any of the last three responses varies from the expected response, make any necessary
adjustments.
21. Type AVGI[ENTER] to display the averaging interval time.
Averaging Interval Time = 00:15
Enter interval (mm:ss):
Figure G-27: Average Interval Time
22. If the averaging interval is 00:15 respond with [ENTER] or 00:15[ENTER]. If the
averaging interval is not 00:15 respond with 00:15[ENTER]. (Averaging is not actually
being used by the ACM, but a zero value will cause the ACM to behave differently when
it first begins logging). The non-zero value 00:15 is the value that the MMP firmware
expects.
23. Run the ROP command again to verify the changes.
G-20