The Auto setting will continuously monitor the voltage output and adjust the gain to
match the signal level. The gain will increase one level when the signal falls below a threshold of
0.5 V and decrease one level when the signal rises above 4 V.
Selection ? T
Set Turbidity Sensor Gain:
<A> Automatic
<1> Fixed 1X
<2> Fixed 3X
<3> Fixed 10X
<4> Fixed 30X
Selection [A] ? a
Figure G-44: Set Turbidity Gain
The sensor data scrolls until the operator selects [CTRL]-[C].
Enter number of measurements to average (1 to 100) ? 20
Press any key to pause/continue display
<X> to change settings, <CTRL>-<C> to exit.
01/07/2005 13:49:04 Turbidity: 1669.97 FTU 3340 mV, 01X Battery: 11.6 V
01/07/2005 13:49:05 Turbidity: 1833.25 FTU 3666 mV, 01X Battery: 11.6 V
01/07/2005 13:49:06 Turbidity: 1774.18 FTU 3548 mV, 01X Battery: 11.6 V
01/07/2005 13:49:07 Turbidity: 1774.75 FTU 3550 mV, 01X Battery: 11.6 V
01/07/2005 13:49:08 Turbidity: 1117.62 FTU 2235 mV, 01X Battery: 11.6 V
01/07/2005 13:49:09 Turbidity: 839.12 FTU 1678 mV, 01X Battery: 11.6 V
Figure G-45: Engineering Data File with Turbidity Voltage and Gain
G-34