Programming a Deployment
When deployment initialization is complete, the Deployment Menu displays. The menu
re-displays after each change.
IMPORTANT
Dedicate the time to thoroughly understand the deployment parameters. These settings guide
and control the operation of the MMP.
ID M| Mooring ID = 001
Start Z| Countdown delay = 12:00:00 [HH:MM:SS]
Schedule I| Profile start interval = 000 12:00:00 [DDD HH:MM:SS]
R| Reference date/time = 01/01/2001 00:00:00
B| Burst Interval = Disabled
N| Profiles per burst = Disabled
P| Paired profiles Disabled
F| Profiles / file set = 1
Stops S| Shallow pressure = 0.0 [dbar]
D| Deep pressure = 6000.0 [dbar]
H| Shallow error = 500.0 [dbar]
E| Deep error = 500.0 [dbar]
T| Profile time limit = 08:00:00 [HH:MM:SS]
C| Stop check interval = 30 [sec]
L| Fluorometer = Enabled Gain Auto Sample/avg 5
O| OBS Turbidity = Enabled Gain Auto Sample/avg 5
Deploy V| Verify and Proceed
Selection ?
Figure G-49: Deployment Initialization (version 3.12)
The parameters are grouped on the display by general function: Mooring ID, Start,
Schedule, Stops and Deploy. The Start parameters, (countdown delay or specific date and time)
control the start time of Profile 0. The MMP executes Profile 0 to move to the bottom stop. The
start delay gives the operator time to launch the MMP and set the mooring before profiling
begins.
A Quick Reference overview of each deployment parameter is provided next. A
laminated version of this information is also included in the Toolkit. More detailed descriptions
of the deployment parameters follow the Quick Reference.
G-40