User manual

Table Of Contents
Detailed Descriptions of MMP Deployment Parameters
A more detailed description of deployment parameter settings follows.
ID M| Mooring ID = 015
R| Reference date/time = 01/01/2001 00:00:00
B| Burst Interval = Disabled
N| Profiles per burst = Disabled
P| Paired profiles Disabled
F| Profiles / file set = 1
Stops S| Shallow pressure = 0.0 [dbar]
D| Deep pressure = 6000.0 [dbar]
H| Shallow error = 500.0 [dbar]
E| Deep error = 500.0 [dbar]
T| Profile time limit = 08:00:00 [HH:MM:SS]
C| Stop check interval = 30 [sec]
L| Fluorometer = Enabled Gain Auto Sample/avg 5
O| OBS Turbidity = Enabled Gain Auto Sample/avg 5
Deploy V| Verify and Proceed
Start Z| Countdown delay = 12:00:00 [HH:MM:SS]
Schedule I| Profile start interval = 000 12:00:00 [DDD HH:MM:SS]
Selection ?
Figure G-50: Deployment Parameters (Firmware version 3.13)
Mooring ID
Mooring ID <M>
Mooring ID provides a three position numeric identifier (001 to 999) for the mooring.
This identifier is stored in the URAO and provides a way to differentiate deployment data when
multiple instruments are deployed. The firmware displays the entry with leading zeros.
Start parameters enable selection of a countdown delay or scheduled start for the
deployment.
Start Parameters
Select Start Parameters <Z>
Select option ‘Z’ to specify the deployment start parameters.
Countdown delay
- A countdown delay in hours, minutes, and seconds can be set
between 00:00:00 and 24:00:00. When the operator commits to the deployment, the countdown
timer begins decrementing. When the alarm is triggered the system wakes up and begins
operations. If an optional transponder is installed, the system pings the transponder at the start of
G-43