User manual

Table Of Contents
accidental launch is less likely (the operator can also terminate the deployment by pressing
[CTRL]-[C]).
IMPORTANT
Before you begin a deployment, ensure that you have an archived copy of the data files stored
on the flash card. Once you begin the deployment, the system will immediately be initialized
for use, which includes reformatting the flash card and erasing any stored data files. After
reformatting is complete, the data files cannot be recovered with disk recovery software.
Default
response
Stores
parameters
One line of
diagnostics
Commit to
deployment
(default is No)
Accept and store schedule selections (Yes/No) [Y] ? y
Storing deployment definition parameters . . . done.
WARNING: If you have not already done so,
REMOVE the flow path CAPS from the CTD NOW.
Failure to remove the caps will prevent proper
CTD operation during the deployment and may
cause permanent damage to the sensor.
After removing the caps, press any key
to continue.
CAUTION: Deployment will erase all data
files stored on the flash card.
System status:
RTC:11/24/2004 15:07:51 WDC: 11/24/2004 15:07:51 10.7 Vb 0 mA
Proceed with the deployment (Yes/No) [N] ?
Figure G-54: Proceed with Deployment
Profile and Deployment Termination Conditions
Profiles and deployments are terminated by any of the following conditions:
shallow or deep pressure limit reached
time limit expired
zero pressure rate inside the shallow or deep error window
operator command
flash card or file system full
low average battery voltage
G-57