Configuring the Firmware to Use a Sea-Bird CTD
The MMP System Configuration menu specifies which sensors are enabled. To enable a
Sea-Bird 41CP or 52MP CTD, complete the following steps:
1. From the Main Menu type ‘c’ and enter the password ‘configure’.
2. Select <2> for the 41CP CTD or <3> for the 52MP CTD and then select ‘Y’ to
enable the sensor.
3. Select [X] to exit and save the entry.
Figure L-3: System Configuration Menu with Sensor Selections
c Password: configure
Synchronizing system configuration files . . . done.
╔═════════════════════════════════╗
System Configuration
╚═════════════════════════════════╝
System Parameters:
<E> Nominal Endurance 1.0 million meters
<I> Inductive Telemetry Disabled
<T> Acoustic Transponder Disabled
<F> File Deletion Disabled
Sensor Suite:
<1> FSI EM CTD Disabled
<2> SeaBird 41CP CTD Disabled
<3> SeaBird 52MP CTD Enabled
<4> FSI 2D ACM Disabled
<5> Nobska MAVS3 ACM Enabled
<6> SeaPoint Fluorometer Disabled (Chlorophyll α)
<7> Wetlabs Fluorometer Disabled
<8> SeaPoint Turbidity Disabled (IR)
Exit:
<X> Save Changes and Exit
Appendix L-2