The 41CP CTD settings can be verified from the Bench Test menu in the MMP firmware.
To display and verify settings, complete the following steps:
21. From the Bench Test option on the Main Menu, select <1> CTD Communications. The
system shows the following display and prompt (S>).
SBE 41CP McLane V 1.0
S>
Figure L-4: 41CP CTD Settings
22. Type ‘ds’ at the prompt to display and verify the CTD settings as shown in Figure L-5.
S>ds
SBE 41CP McLane V 1.0 SERIAL NO. 1001
stop profile when pressure is less than = -100.0 decibars
automatic bin averaging when p < -100.0 disabled
number of samples = 0
number of bins = 0
top bin interval = 10
top bin size = 10
top bin max = 100
middle bin interval = 50
middle bin size = 50
middle bin max = 1000
bottom bin interval = 100
bottom bin size = 100
do not include two transition bins
S>
Figure L-5: Verifying 41CP CTD Settings
23. At the next S>prompt, type [CTRL]-[C] to power off the CTD.
24. Type <CR> to return to the Bench Test Menu.