Verifying 52MP CTD Settings
The 52MP CTD settings can be verified from the Bench Test menu in the MMP
firmware. To display and verify settings, complete the following steps:
5. From the Bench Test option on the Main Menu, select <1> CTD Communications. The
system shows the following display and a prompt (S>).
SBE 52 MP 1.0
S>
Figure L-6: 52MP CTD Settings
6. Type ‘ds’ at the prompt to display and verify CTD settings as shown in Figure L-7.
Sending command ds
SBE 52 MP 1.0 SERIAL NO. 0003
stop profile when pressure is less than = -100.0 decibars
automatic bin averaging when p < -100.0 disabled
number of samples = 0
number of bins = 0
top bin interval = 10
top bin size = 10
top bin max = 100
middle bin interval = 50
middle bin size = 50
middle bin max = 1000
bottom bin interval = 100
bottom bin size = 100
do not include two transition bins
oxygen frequency multiplier = 0.25
S>
Figure L-7: Verifying 52MP CTD Settings
7. At the next S>prompt, type [CTRL]-[C] to power off the CTD.
8. Type <CR> to return to the Bench Test Menu.
Appendix L-4