27. Type ‘Get_All’ to display Optode settings as shown in Figure M-6. Verify the following
settings: Interval = 30 and Output = 100.
Get_All
Protect 3830 688 0
PhaseCoef 3830 688 -8.323242E+00 1.183118E+00 0.000000E+00 0.000000E+00
TempCoef 3830 688 2.609516E+01 -3.167202E-02 2.981936E-06 -4.364193E-09
FoilNo 3830 688 4804
C0Coef 3830 688 3.172420E+03 -1.072609E+02 2.133159E+00 -1.792340E-02
C1Coef 3830 688 -1.739810E+02 5.102620E+00 -9.857580E-02 8.022400E-04
C2Coef 3830 688 3.956000E+00 -9.810400E-02 1.853460E-03 -1.425360E-05
C3Coef 3830 688 -4.263370E-02 8.788200E-04 -1.644089E-05 1.137090E-07
C4Coef 3830 688 1.768690E-04 -2.955020E-06 5.550390E-08 -3.084930E-10
Salinity 3830 688 0.000000E+00
CalAirPhase 3830 688 2.985921E+01
CalAirTemp 3830 688 1.987216E+01
CalAirPressure 3830 688 1.014400E+03
CalZeroPhase 3830 688 6.518473E+01
CalZeroTemp 3830 688 2.047768E+01
Interval 3830 688 30
AnCoef 3830 688 0.000000E+00 1.000000E+00
Output 3830 688 100
SoftwareVersion 3830 688 3
Interval
Output
SoftwareBuild 3830 688 7
Figure M-6: Verifying Optode Settings
28. Refer to the Aanderaa documentation for other valid direct commands.
29. Type [CTRL]-[C] to return to the Bench Tests menu.