Collecting Data with the Aanderaa Optode
Optode data is logged in the Engineering File as shown in Figure M-8. When the Optode
is disabled or switched off (as during ramp up), the columns contain ‘99.90’.
Figure M-8: Engineering Data with Aanderaa Optode Oxygen and Temperature Data
IMPORTANT
The Optode data collection interval is the same as the MMP firmware ‘stop check interval’
setting.
Profile 0
Sensors were turned on at 03/14/2006 08:38:18
Vehicle began profiling at 03/14/2006 08:40:15
Date Time [mA] [V] [dbar] Fluor[mV] Gain Turb[mV] Gain Oxygen[uM] Optode Temp[C]
03/14/2006 08:40:15 -3 11.4 0.000 99.00 99 99.00 99 99.90 99.90
03/14/2006 08:40:30 17 11.4 0.000 99.00 99 99.00 99 99.90 99.90
03/14/2006 08:40:45 17 11.4 0.000 99.00 99 99.00 99 99.90 99.90
03/14/2006 08:40:46 17 11.4 0.000 99.00 99 99.00 99 279.57 19.37
03/14/2006 08:41:01 16 11.5 15.000 99.00 99 99.00 99 280.02 19.38
03/14/2006 08:41:16 16 11.5 30.000 99.00 99 99.00 99 280.05 19.39
03/14/2006 08:41:31 16 11.5 45.000 99.00 99 99.00 99 280.09 19.40
03/14/2006 08:41:46 16 11.5 59.000 99.00 99 99.00 99 280.16 19.41
03/14/2006 08:42:01 16 11.5 74.000 99.00 99 99.00 99 280.02 19.42
03/14/2006 08:42:16 16 11.5 90.000 99.00 99 99.00 99 280.07 19.44
03/14/2006 08:42:31 16 11.5 105.000 99.00 99 99.00 99 279.95 19.45
03/14/2006 08:42:46 16 11.5 120.000 99.00 99 99.00 99 279.80 19.46
03/14/2006 08:43:01 16 11.5 134.000 99.00 99 99.00 99 279.83 19.47
03/14/2006 08:43:16 16 11.5 150.000 99.00 99 99.00 99 279.86 19.48
03/14/2006 08:43:31 16 11.5 165.000 99.00 99 99.00 99 279.69 19.50