At each MMP ‘stop check interval’ the Optode is powered on, data is collected and the
Optode is powered off. As a result, the ‘stop check interval’ setting affects overall battery
endurance.
The examples below show how the ‘stop check interval’ impacts battery endurance
(based on a 1000m profile). The Optode uses .0205 mAh per data point (average). The MMP
battery endurance is approximately 240,000 mAh (240 Ahr).
Sample Endurance Calculations
Stop Check Interval Setting
(seconds)
Data Points Per Profile Optode Battery
Requirements
(% of battery)
4 1000 7%
15 267 2%
30 133 1%
Setting the ‘stop check interval’ at 4 seconds requires 7% of the battery. A less frequent
stop, such as 30 seconds, requires 1% of the battery.