User manual

Table Of Contents
IMPORTANT
When positioning the ACM, place the MMP right side down (with the McLane label visible on
the drive motor as in Figure 2-13). Laying the MMP on the right side prevents the ACM sting
from pivoting down when the cap screws securing the hinged mounting bracket are removed.
Be sure to re-secure the screws for the deployment. The hinge is attached so that the
sting will rest in position when the MMP is lying on its right side.
NOTE
The mooring cable must be located on the center line to properly align the MMP into the flow.
An off-center location introduces a left-right bias in the flow measurement because the sting is
not symmetrically aligned with the flow streamlines around the body of the MMP.
After orienting and attaching the sting, the cable should be secured to the hinged bracket
with a plastic tie-wrap as shown in Figure 2-14 to minimize contact with the mooring cable
running inside the bracket. There is a hole in the corner of the bracket to attach the tie-wrap.
Lay the oil-filled cable through the slot in the left side skin and along the grooves in the
frame ribs and spacer posts and thread through the narrow slot in the hinged bracket. Do not
force the cable immediately adjacent to the sting base plate through the slot. Release enough
cable from the interior of the MMP so that the sting base plate is 15cm (6”) from the slot.
Squeeze the cable gently with two fingers, aligning the interior wires so that they do not cross
over each other, and slide the cable through the slot.
IMPORTANT
The oil filled cable connecting the ACM sting to the ACM electronics housing is fragile. Do
not pull on or introduce a tight bend in the cable and avoid crimping the outer tube. Damage to
the cable must be repaired by the manufacturer. Be gentle.
Excess cable can be coiled loosely near the ACM electronics housing.
2-11