User manual

Table Of Contents
Sea-Bird CTD
The screens and descriptions shown next refer to the Sea-Bird 41CP and 52MP CTD (the
52MP CTD may have an optional Dissolved Oxygen Sensor).
NOTE
More information about Sea-Bird CTD sensors is provided in this User Manual in the “Optional
Sea-Bird CTD Sensors” chapter.
Verifying 41CP CTD Settings
To use the CTD pass-through utility complete the following steps:
1. From the Bench Test option on the Main Menu, select <1> CTD Communications..
SBE 41CP McLane V 1.0
S>
Figure 7-23: 41CP CTD Settings
2. Type ‘ds’ at the prompt to display and verify the 41CP CTD settings (see Figure 7-24).
S>ds
SBE 41CP McLane V 1.0 SERIAL NO. 1001
stop profile when pressure is less than = -100.0 decibars
automatic bin averaging when p < -100.0 disabled
number of samples = 0
number of bins = 0
top bin interval = 10
top bin size = 10
top bin max = 100
middle bin interval = 50
middle bin size = 50
middle bin max = 1000
bottom bin interval = 100
bottom bin size = 100
do not include two transition bins
S>
Figure 7-24: Verifying 41CP CTD Settings
7-17