IMPORTANT
To terminate the communication session and power down the sensor, type [CTRL]-[C] at
anytime, regardless of the current operational mode.
The FSI ACM has standard settings for use with the MMP. These settings can be
confirmed before beginning a deployment, while in direct communication with the ACM. The
firmware will not automatically program the standard settings.
To verify and program FSI ACM settings, complete the following steps:
1. From the Bench Test option on the Main Menu, select <5> ACM Communications.
2. Type ***O [ENTER] to place the ACM in open mode.
Open mode
cmd followed
by [ENTER]
***O
OPEN MODE
Figure 7-31: ACM in Open Mode
3. Type RDM [ENTER]. The following should display:
TX=ON, TY=ON, HX=ON, HY=ON, HZ=ON, VPATH=ON, 18
Figure 7-32: ACM Settings
4. The response to Read Data Message (RDM) indicates which features are to be logged
(tilt, compass, path velocities). If the response differs from Figure 7-32, disable settings
that should be ‘Off’ by typing
NAME=OFF[ENTER], e.g., ATIME=OFF[ENTER] and enable
settings that should be ‘On’ by typing
NAME=ON[ENTER], e.g., HZ=ON[ENTER]. The number
at the end of the RDM display is the number of bytes per scan in the compressed binary
format.
NAME[ENTER] displays the status of that particular feature.
5. Type ROP [ENTER]. The following should display:
Continuous clear
address op clear
Logging Ops Set
RLD checksum output cleared
Figure 7-33: ACM Operational Parameters
7-21