If the response to Read Operational Parameters (ROP), varies from the expected one,
enter commands from the list below to correct the settings:
Command Result
CCOP[ENTER]
Clears continuous
CAOP[ENTER]
Clears address operation
SLOG[ENTER]
Sets logging ops (the operator cannot control the checksum output)
TILT[ENTER]
TILT= on
(Enables tilt)
COMP[ENTER]
COMP = on
(Enables compass)
NRML[ENTER]
NRML = on
(Sets compass normalization)
6. If any of the last three responses varies from the expected response, make any necessary
adjustments.
7. Type AVGI[ENTER] to display the averaging interval time.
Averaging Interval Time = 00:15
Enter interval (mm:ss):
Figure 7-34: Average Interval Time
8. If the averaging interval is 00:15 respond with [ENTER] or 00:15[ENTER]. If the
averaging interval is not 00:15 respond with 00:15[ENTER]. (Averaging is not actually
being used by the ACM, but a zero value will cause the ACM to behave differently when
it first begins logging). The MMP firmware expects the non-zero value 00:15.
9. Run the ROP command again to verify the changes.
10. Once all of the settings are correct, type ***E[ENTER] to save settings in the ACM
EEPROM.
11. Wait a few seconds for the ACM to add a carriage return, indicating that the storage
operation is complete.
12. Press [ENTER] to return the ACM to Open Mode.
13. Press [CTRL]-[C] to terminate the session and power down the ACM.
14. Re-establish communications and check that the settings are now correct on power-up.
A sample screen for verifying standard settings is shown next.
7-22