Preinstallation Manual Verdi™ V-2/V-5/V-6 Diode-Pumped Lasers 5100 Patrick Henry Drive Santa Clara, CA 95054
Verdi V-2/V-5/V-6 Laser Preinstallation Manual This document is copyrighted with all rights reserved. Under the copyright laws, this document may not be copied in whole or in part or reproduced in any other media without the express written permission of Coherent, Inc. Permitted copies must carry the same proprietary and copyright notices as were affixed to the original.
Table of Contents TABLE OF CONTENTS Preface ............................................................................................................................ iv U.S. Export Control Laws Compliance .......................................................................... iv Symbols Used in This Manual and on the Laser System .................................................v Introduction..............................................................................................................
Verdi V-2/V-5/V-6 Laser Preinstallation Manual Preface This manual contains user information for the Verdi™ V-2/V-5/V-6 Diode-Pumped Lasers. Read the Operator’s Manual carefully before operating the laser for the first time. Special attention should be given to the material in Section One, Laser Safety, that describes the safety features built into the laser. Use of controls or adjustments or performance of procedures other than those specified in this manual may result in hazardous radiation exposure.
Preface Symbols Used in This Manual and on the Laser System This symbol is intended to alert the operator to the danger of exposure to hazardous visible laser radiation. This symbol is intended to emphasize the presence of important operating instructions.
Verdi V-2/V-5/V-6 Laser Preinstallation Manual vi
Verdi V-2/V-5/V-6 Laser Preinstallation Manual Introduction The purpose of this document is to assist the user in establishing a suitable location and operating environment for optimum Verdi V-2/V-5/V-6 performance. Receiving and Inspection Inspect shipping containers for signs of rough handling or damage. Indicate any such signs on the bill of lading.
Verdi V-2/V-5/V-6 Laser Preinstallation Manual 2 3 5 7 10 6 9 8 1 4 1. Riser/heat sink installation hardware 6. Interlock-disassembled 2. Operator’s manual 7. Fuses and instructions 3. AC power cord 8. 4. Power supply key FAP-I installation hardware, including shorting clips, end caps, etc. 5. Installation kit case 9. Laser head/heat sink mounting hardware 10. Dowel pins – laser head/heat sink Figure 1-1.
Verdi V-2/V-5/V-6 Laser Preinstallation Manual The laser head is shipped with a protective plastic plate installed over the baseplate. Leave the plastic plate in place until laser head is ready for installation. 1. Remove the packing material from the riser/heat sink (Figure 1-2). 2. Insert two dowel pins (Figure 1-1 , item 12) into the riser/heat sink dowel holes (Figure 1-2, item 1), ball-end up. Avoid damage to the laser head baseplate, which is milled to a flatness of ± 0.1 mm.
Verdi V-2/V-5/V-6 Laser Preinstallation Manual 1 2 3 4 1. Dowel holes (2) 3. 2. Cooling water inlet and outlet (if included) 4. Laser head clamp screw holes (4) Optical table clamp mounts (4) Figure 1-2. Riser/Heatsink 1 4 3 2 1. Laser head 3. Protective material 2. Protective plastic cover retaining screws (3) 4. Protective plastic cover Figure 1-3.
Verdi V-2/V-5/V-6 Laser Preinstallation Manual DOWEL PIN HOLES Figure 1-4. Laser Head Base – Protective Cover Removed Installing the Power Supply 6. Proceed to the paragraph titled, “Installing the Power Supply”. The riser/heat sink clamps and chiller connection will be installed during the final installation procedures. 1. Move the power supply to an accessible-friendly location, preferably away from heat producing sources.
Verdi V-2/V-5/V-6 Laser Preinstallation Manual Each customer must inspect the power cord and install the proper connector if necessary. The connector should be installed in a properly grounded outlet with a maximum of 10 Ampere service @ 110 V and 11 Ampere service @ 220 V for proper overcurrent and earth fault protection. 4. Final Installation and Initial Turn-On Procedures Proceed to the paragraph titled, “Final Installation and Initial Turn-On Procedures”.
Verdi V-2/V-5/V-6 Laser Preinstallation Manual 7. External Interlock Perform the procedure in paragraph titled, “Turn-on (Cold-Start)” in the Operator’s Manual. The system will not operate with the interlock open. An interlock connector is located on the power supply rear panel. The interlock status is monitored by the CPU. If the interlock is open, the message, “External Interlock Fault” will be displayed on the power supply front panel.
Verdi V-2/V-5/V-6 Laser Preinstallation Manual 5. The interlock circuit in Verdi operates from a +/-12 VDC circuit. Its current is limited to around 1 mA. For these ratings, a “dry-circuit” type switch will give the most reliable operation. UMBILICAL AIR COOLED POWER SUPPLY LASER HEAD FACILITY POWER WATER LINES (if using Water Cooling) CHILLER FACILITY POWER Figure 1-6. Interconnect Diagram Table 1-1.
Verdi V-2/V-5/V-6 Laser Preinstallation Manual Table 1-2. Environmental Requirements PARAMETER REQUIREMENT Operating temperature 15 to 35°C (59 to 95°F) Relative humidity 5 to 95% (non condensing) Altitude Sea level to 10,000 feet Dimensions Figure 1-7.
Verdi V-2/V-5/V-6 Laser Preinstallation Manual Figure 1-8.
Verdi V-2/V-5/V-6 Laser Preinstallation Manual 11
Verdi V-2/V-5/V-6 Lasers Preinstallation Manual © Coherent, Inc. 09/2005, Printed in the U.S.A. Part No.