5. The interlock circuit in Verdi operates from a +/-12 VDC
circuit. Its current is limited to around 1 mA. For these ratings,
a “dry-circuit” type switch will give the most reliable opera-
tion.
Figure 1-6. Interconnect Diagram
Table 1-1. Utility Requirements
PARAMETER REQUIREMENT
Power Requirements 100 to 240 VAC
[1][2]
Maximum Current 6.5 Amps @ 100 VAC
7 Amps @ 220 VAC
Cooling:
Laser head:
Power supply:
Riser/heat sink
[3]
Air-cooled with ambient air
Note: All specifications and requirements are subject to change without notice.
[1] The power supply is autoranging and will accommodate the full range of input voltages without hardware
changes.
[2] The electrical service should have a main power disconnect switch located in close proximity to the laser. The
main power disconnect switch shall be clearly marked as the disconnecting device for the laser, and shall be
within easy reach of the operator.
[3] The passive riser/heat sink must be mounted on a conductive surface capable of maintaining the laser head base-
plate at less than 55°C. An optional riser/heat sink and chiller or chill plate can also be used. Refer to the para-
graph titled, Installing the Laser Head, for additional information on removing waste heat from the laser head.
LASER HEAD
AIR COOLED
POWER SUPPLY
UMBILICAL
FACILITY
POWER
WATER LINES (if using Water Cooling)
CHILLER
FACILITY
POWER