1
Introduction
The purpose of this document is to assist the user in establishing a
suitable location and operating environment for optimum Verdi
V-2/V-5/V-6 performance.
Receiving and
Inspection
Inspect shipping containers for signs of rough handling or damage.
Indicate any such signs on the bill of lading. Report any damage
immediately to the shipping carrier, and to Coherent Order Admin-
istration Department (800-438-6323) or to an authorized Coherent
representative.
Retain shipping containers. The containers will be required if
the system is returned to the factory for service. The containers
may also be needed to support a shipping damage claim.
Coherent recommends that three people unpack and transport the
Verdi V-2/V-5/V-6. The power supply weighs70 pounds and should
be lifted by two people while the laser head, which weighs 22.2
pounds, can be lifted by the third person. The laser head and power
supply are connected by the umbilical.
Installation Kit
The installation kit contains the necessary hardware for Verdi instal-
lation. See Figure 1-1 for details.
Installing the
Laser Head
The laser head is shipped with either a passive or a water-cooled
riser/heat sink. This riser/heat sink maintains the Verdi output beam
at a suitable height for pumping a laser such as the Coherent Mira.
A passive riser/heat sink when mounted on a thermally conductive
surface (such as an optical table) provides sufficient heat dissipation
for the laser head under normal ambient conditions. The temperature
of the laser head baseplate is monitored by the Verdi power supply
and can be displayed on the power supply front panel. Power dissi-
pation through the baseplate is typically 20 to 25 Watts (30 Watts
maximum). The Verdi will shut down if the baseplate temperature
exceeds 55°C.
An optional water-cooled riser/heat sink is available for use with a
chiller.