Instruction manual
Hallett 15xs Instruction Manual 
Released on 20-Sept-12 – H800003P  Page 28  UV Pure Technologies Inc. © 2012 
SETUP MENU 
Setup Menu 
Hallett 15xs
Language  
English
Language  
English/French
Remote StartStop 
Disabled
Remote StartStop 
Disable/Enable
Shutoff Valve  
Installed
Shutoff Valve  
Not Installed / Installed
Unit 
Configuration
Enter Password 
######
Remote Device 
BIND #
Bind Remote 
Device 
Remote Device 
BOUND
ALARMS & TROUBLESHOOTING MENU 
Alarms and 
Troubleshooting
Lamp Alarm -  
Low UV Intensity
Lamps need to 
burn in
Continue to 
operate lamps
UV Air Temp. 
High
Check if filter 
dirty, fan fail
Check if room 
temp within spec
UV Air Temp. 
Low
Check if room 
temp within spec
Switch lamps & 
examine values
Only one lamp 
may be low
Sensor may 
have drifted
New sensor or 
calibration reqd
Lamps require 
replacement
Install 2 lamps 
& reset counter
Front Panel ajar
Fully Close  
front panel
Lamps not 
striking
Front Panel ajar
Fully Close  
front panel
Check lamps 
or ballast
Lamp Lifetime  
exceeded
Lamps require 
replacement
Install 2 lamps 
& reset counter
End of Lamp Life 
Approaching
Lamp cycling 
too often
Reduce the # of 
lamp starts/day
ALARM
WARNING
ALARM
ALARM
WARNING
This column represents all possible 
reasons for the fault.