Cleaning/Removing the Quartz Sleeve 
Note that numbers in parentheses refer to Figure 4B. 
Step 1: Use a bottle cleaning brush with a long handle to scrub the inside of the 
quartz sleeve. Scrub and flush it with water repeatedly to clean the quartz. Use a 
squeeze bottle to apply water or solution to the quartz sleeve to keep the area 
tidy. Note: Keep the rest of the unit free from moisture. Examine the quartz. 
Step 2: If the quartz is still dirty, use a scale remover such as CLR or Lime Away 
and apply it to the inside of the quartz sleeve. Citric acid, available at a drug 
store, can also be used. Always flush with clean water afterwards. 
Step 3: Once the quartz is clean, reassemble the unit (see Figure 4E). Replace 
any seals that appear to have been damaged. 
Step 4: If the quartz is still not clean, it requires replacement. This is done by 
removing the bottom quartz seal assembly (see Figure 4D). 
Step 5: Remove the bottom cover and then the bottom quartz seal assembly by 
pulling the plastic tabs of the cover out and down and then undoing the four nuts 
in an alternating pattern (top left, bottom right, bottom left, then top right). Support 
the quartz sleeve as you remove this item. 
Step 6: Remove the quartz sleeve by removing the top and bottom O rings. Lift 
the quartz sleeve out of the unit. 
Step 7: Install the new quartz sleeve into the unit and center it vertically. Be 
careful not to chip the ends.  You may open the front of the unit to assist you in 
aligning the quartz sleeve if necessary. Support the quartz for the next two 
actions. 
Step 8: Replace the top and bottom O rings, keeping the quartz centered 
vertically in the unit. 
Step 9: Replace the bottom quartz seal assembly (see Fig. 4D) by installing the 
four nuts in an alternating pattern (top left, bottom right, bottom left, then top 
right) Ensure that the connection port is at the backside of the unit 
Reassembling the Unit 
Note that numbers in parentheses refer to Figure 4B. 
Step 1: Replace the wiper assembly (12) carefully in the quartz sleeve. The hole 
on the end of the shaft must be up. Turn the wiper assembly counter-clockwise 
(looking from the top) as it is being inserted into the quartz – this will make the 
task easier and align the wiper blades properly. Wetting the inside of the quartz 
with water will also make the task easier. 
Step 2: Ensure the quartz O-ring seal is in place around the quartz sleeve and 
then replace the top quartz seal assembly by aligning the shaft adapter with the 
wiper shaft (see Figure 4E). Rotate wiper shaft to suit. 