User Manual
Specications 8
30 Series Operation Manual
PRESSURE RANGES, DISPLAY SCALES, & RESOLUTION
PSI bar kPa Overpressure PSI inch H
2
O inch Hg mmHg mmH
2
O cmH
2
O kg/cm
2
bar mbar kPa MPa
16 1 100 6.5 x 0.001 0.01 0.001 0.01 1 0.0001 0.0001 0.1 0.01
36 2 200 3.0 x 0.001 0.01 0.001 0.01 1 0.0001 0.0001 0.1 0.01
300 20 2000 2.0 x 0.01 0.1 0.01 0.1 1 0.001 0.001 0.1
1500 100 10 000 2.0 x 0.1 0.01 0.01 1 0.001
3000 200 20 000 1.5 x 0.1 0.01 0.01 1 0.001
5000 300 30 000 1.5 x 0.1 0.01 0.01 1 0.001
•
PSI versions with ranges of 36 PSI and lower include all possible scales except bar.
•
bar versions with ranges of 2 bar and lower include all possible scales except inch Hg, cmH2O, and MPa.
•
kPa versions are restricted to kPa, MPa, bar and/or millibar, depending on pressure range.
MODEL NUMBERING SYSTEM
Available Units
Intrinsic
Safety
Model Sensors
1st Pressure
Range Prex
2nd Pressure
Range Prex
Pressure
Unit
Manifold
-
IS 3
- / -
Standard ...........omit Single Sensor ...1 PSI ....PSI Standard ........omit
Metric units only
...SI Dual S
ensor
.....3 bar ...BAR S
tainless Steel
...SPSS
kPa
...KPA
SAMPLE PART NUMBERS
IS31-36PSI ...............Calibrator with one sensor (36 PSI)
IS33-36/3000PSI ..........Calibrator with two sensors (36 PSI and 3000 PSI)
IS33-16/1500PSI-SPSS ...Calibrator with two sensors (16 PSI and 1500 PSI) with two stainless steel manifolds
SERIAL NUMBERING SYSTEM
The serial number of your 30 Series is located on the rear of the product. Serial numbers consist of 10 numbers, with the left-most digit of the second
grouping representing the year of manufacturing. For example 043468 was manufactured in 2010.