User Manual
Specications 7
30 Series Operation Manual
Power
Battery . . . . . . . . . . . . . . . . . . . . . . . . . . One Alkaline or Lithium 9V battery
X
Approved Batteries
ATEX/IECEx: The 30 Series is Intrinsically Safe only if powered by one of the following battery types:
Approved Battery Type Ta= Marking
Duracell Alkaline 9V, MN 1604
0 to 50° C
Ex ia IIC T4 Gb
Energizer Alkaline 9V, 522
Energizer Lithium 9V, L522
Varta High Energy Alkaline 9V, 4922 0 to 45° C
Many other battery types and models have been tested but failed to meet the requirements for potentially explosive atmospheres—do not assume other
models are equivalent.
Do not mix battery types or manufacturers.
CSA: Substitution of components may impair intrinsic safety.
CSA: La substitution de composants peut compromettre la sécurité intrinsèque.
CSA: The 30 Series is Intrinsically Safe only if powered by one of the following battery types:
CSA: Le 30 Series est un système à sécurité Intrinsèque seulement s’il est alimenté par un des Piles de type suivant:
Approved Battery Type
(Approuvé Type de Batterie)
Marking
(Marquage)
Duracell Alkaline 9V, MN 1604
Class I, Division 1, A, B, C, D T4
(Classe I, Division 1, A, B, C, D T4)
Energizer Alkaline 9V, 522
Energizer Lithium 9V, L522
Varta High Energy Alkaline 9V, 4922
Class I, Division 1, A, B, C, D T3C
(Classe I, Division 1, A, B, C, D T3C)
Battery Life . . . . . . . . . . . . . . . . . . . . . . 90 hours typical
Low Battery Indicator . . . . . . . . . . . . Flashing battery icon
Connections
Pressure Fitting . . . . . . . . . . . . . . . . . . ⁄" female NPT
Milliamperes . . . . . . . . . . . . . . . . . . . . 4mm jacks, 19mm (0.75”) spacing
Communication . . . . . . . . . . . . . . . . . micro USB
!
WARNING: Do not use the micro USB serial interface in hazardous locations.