APM CPF Series Operation Manual for Advanced Pressure Module
Contents Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Overview 1 Overview INTRODUCTION The Crystal APM CPF series of pressure modules allow you to add pressure measurement capability to your JOFRA process calibrator. The APM uses the same reliable, high accuracy, digital temperature compensated technology found in the nVision, housed in a new rugged enclosure with an integrated cable to connect to your JOFRA calibrator. In addition, you can connect the APM to a hand-held pressure calibrator to increase the range of that calibrator.
Operation 2 Operation ADVANCED PRESSURE MODULE (APM) INSTRUCTIONS Pressure Connection Crystal CPF System: Medium Pressure Female (MPF) (1/4” medium pressure tube system with 7/16-20 threads). See our CPF Brochure for further information. CPF o-ring size and material: AS568A-012, Viton 80 durometer (P/N 3981). For most applications CPF Fittings can be hand tightened for use up to 10 000 psi / 700 bar / 70 MPa (no tools required).
Specifications 3 Specifications Accuracy (Gauge) Modules psi bar kPa / MPa Module Range 0 - 30% Gauge Full Scale 30 - 110% Gauge Full Scale psi 30 100 300 1000 3000 10000 15000 ± (% Full Scale) 0.0075% 0.0075% 0.0075% 0.015% 0.015% 0.03% 0.03% ± (% of Reading) 0.025% 0.025% 0.025% 0.05% 0.05% 0.1% 0.1% bar 3 10 30 100 300 700 1000 ± (% Full Scale) 0.0075% 0.0075% 0.0075% 0.015% 0.015% 0.03% 0.03% ± (% of Reading) 0.025% 0.025% 0.025% 0.05% 0.05% 0.1% 0.
Specifications 4 Accuracies, Ranges, and Resolutions psi bar kPa/MPa 30PSI 0.025% of Reading modules 3BAR 300KPA 100PSI 10BAR 1MPA 300PSI 30BAR 3MPA 1KPSI 0.05% of Reading modules 100BAR 10MPA 3KPSI 300BAR 30MPA 10KPSI 0.1% of Reading modules 700BAR 70MPA 15KPSI* 1000BAR* 100MPA* Overpressure psi kg/cm2 inHg inH20 mmHg mmH20 kPa bar mbar 3.0 x 3.0 x 3.0 x 2.0 x 2.0 x 2.0 x 2.0 x 2.0 x 2.0 x 2.0 x 2.0 x 2.0 x 1.5 x 1.5 x 1.5 x 1.5 x 1.5 x 1.5 x 1.3 x 1.3 x 1.3 x 0.001 0.001 0.
Specifications 5 Pressure Conversions Pressure conversion factors will vary, and be dependent on the JOFRA device connected to the APM. Refer to the documentation for your particular JOFRA device for its specifications. Media Compatibility Liquids and gases compatible with sensor and CPF fitting system.
Communications 6 Communications The APM interface uses a command – query – response protocol. Using this protocol the user can access the pressure measured by the module, as well as control and monitor the status of all the essential functions of the module. INTERFACE DESCRIPTION The APM can be remotely controlled using a PC terminal or computer program running the module in an automated system. A USB-based serial port connection is required for remote operation.
Communications 7 Command Processing The data entered into the module is processed as follows: • • • • • Data is taken as 7-bit ASCII. Characters are discarded if their decimal equivalent is less than 32 (space), except 10 (LF) and 13 (CR). The most significant data bit is ignored. Upper or lower case is acceptable. More than one command (Compound Commands) may be entered on one line. Commands are separated by semi-colons.
Communications 8 In order to avoid interruptions in the command execution process: • • • • Turn on software ( XON / XOFF ) flow control on the PC side. Avoid long command lines (lines more than 96 characters). Send the *CLR command before starting the communication procedure. Check the error buffer sending the FAULT? command after the communication procedure is finished. REMOTE COMMANDS AND ERROR CODES The following tables list all the commands, and their descriptions, accepted by the APM.
Communications 9 Parameter Units Units Meaning PSI Pressure in pounds per square inch. KG/CM2 or KGCM2 Pressure in kilograms per square centimeter. INHG Pressure in inches of mercury at 0°C. INH2O4C Pressure in inches of water at 4°C. INH2O20C Pressure in inches of water at 20°C. INH2O60F Pressure in inches of water at 60°F. INH2O68F Pressure in inches of water at 68°F. MMHG Pressure in millimeters of mercury at 0°C. MMH2O4C Pressure in millimeters of water at 4°C.
Communications 10 ENTERING COMMANDS Commands for the APM may be entered in upper or lower case. At least one space must be entered between the command and parameter. All other spaces are optional. Almost all commands for the APM are sequential, and any overlapped commands will be indicated as such. This section will briefly explain each of the commands and describe their general use.
APMCAL 11 APMCAL CALIBRATION SOFTWARE APMCAL provides an easy way to calibrate your APM CPF Series Pressure Module. With a USB to serial cable, a computer running Windows, an APM module, and the APMCAL software program, you can: • • • • Calibrate the APM with an easy-to-use Userspan Wizard. View and change the calibration date. View module details including model number, serial number, and firmware version. Update the firmware version of the APM.
Certifications 12 Certifications CE Statement We declare that the APM is in accordance with the Electromagnetic Compatibility Directive, and the Pressure Equipment Directive per our declaration(s). CE PED EC Declaration of Conformity EMC EC Declaration of Conformity DECLARATION OF CONFORMITY DECLARATION OF CONFORMITY According to ISO/IEC 17050-1:2004 According to ISO/IEC 17050-1:2010 Manufacturer’s Name: Crystal Engineering Corporation An AMETEK Inc.
Support 13 Support CONTACT US USA Crystal Engineering • California* Tel +1 (800) 444 1850 Fax +1 (805) 595 5466 crystal@ametek.com Mansfield & Green • Florida Tel +1 (800) 527 9999 cal.info@ametek.com United Kingdom Tel +44 (0)1243 833 302 jofra@ametek.co.uk Germany Tel +49 (0)2159 9136 510 info.mct-de@ametek.de India Tel +91 22 2836 4750 jofra@ametek.com France Tel +33 (0)1 30 68 89 40 general.lloyd-instruments@ametek.fr Denmark Tel +45 4816 8000 jofra@ametek.
© 2014 Crystal Engineering Corporation 708 Fiero Lane, Suite 9, San Luis Obispo, California 93401-8701 5274.